When you select a data collection rate, observe the following guidelines:
•
In general, each peak should be defined by at least 20 data points.
For chromatograms with co-eluting peaks or low signal-to-noise
ratios, 40 data points per peak are recommended.
•
If all peaks are relatively wide, select a lower data collection rate (for
example, 1.0 Hz). This saves disk space and allows for a faster display
of data in the Chromeleon software.
•
If the data collection rate is too low, the start points and end points
of peaks will not be determined accurately.
•
If any peaks of interest are less than a few seconds, select a higher
data collection rate (10.0 Hz, for example). If the data collection rate
is too high, data files may need more disk space and post-run
analyses may require more processing time.
•
Consider also the filter constant when setting the data collection
rate. Set the two parameters together in order to optimize the
amount of data points collected, and reduce short-term noise, while
still maintaining peak height, symmetry, and resolution. See below.
Filter Constant
The filter constant is applied to the output current to electronically
reduce the baseline noise during the analysis.
When setting a filter constant, observe the following:
•
A small filter constant, such as 0.2 seconds, removes little baseline
noise.
•
A large filter constant, such as 10.0 seconds, smoothes the baseline
significantly.
Summary of Contents for Vanquish VF-D20
Page 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Page 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Page 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Page 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Page 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Page 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Page 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...