3
•
Detector Overview
Vanquish Charged Aerosol Detectors Operating Manual
Page 31
3.2
Operating Principle
The detector is designed for charged aerosol detection with full
compatibility in HPLC and UHPLC applications. In charged aerosol
detection, an aerosol of dried analyte particles is formed after
nebulization of column eluate. Electrical charge is imparted to these
dried particles. The detector then measures the charge. Measured
charge is proportional to the amount of analyte in the sample.
The following picture shows the interior components of the detector,
and illustrates how the detector operates:
Figure 1: Operating principle of the detector
No. Description
No. Description
1
Eluate from column
8
Mixing chamber
2
Nebulizer
9
Charger gas inlet
3
Gas inlet
10
Corona charger
4
Gas filters
11
Ion trap
5
Spray chamber
12
Electrometer
6
Drain pump
13
Electrometer board
7
Evaporation tube
14
Gas exhaust
Eluate from the column enters the detector (no. 1) and flows to the
nebulizer (no. 2). Pressurized gas streams through the gas inlet (no. 3)
and passes through the charcoal gas filter and then the HEPA gas filter
(no. 4). In the nebulizer (no. 2), the filtered gas and the eluate merge at
the nebulizer tip to create an aerosol within the spray chamber (no. 5).
Summary of Contents for Vanquish VF-D20
Page 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Page 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Page 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Page 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Page 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Page 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Page 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...