Mind the correct orientation of the nebulizer: The gripping sides
of the nebulizer are on the top and bottom side of the handle.
The protruding sides for attachment are on the left and right
side of the handle. See the picture below.
Figure 14: Correct orientation of the nebulizer for installation
3. Push the nebulizer into the opening until the spray pipe and sealing
bushing are completely in the opening. The nebulizer handle must
lie flush against the nebulizer opening.
Figure 15: Completely inserting the nebulizer
No. Description
No. Description
1
Opening for the nebulizer on the
interior panel of the detector
2
Spray pipe
3
Sealing ring in the sealing bushing
No.
Description
1
Opening in the middle of
the nebulizer opening
2
Gripping sides on top and
bottom handle side
3
Protruding sides on left
and right handle side