6
•
Operation
Vanquish Charged Aerosol Detectors Operating Manual
Page 99
6.5.3
Turning On the Detector Gas Flow
Follow the instructions in this section to turn on the detector gas flow.
CAUTION—Escaping Gas
Gas can escape from open gas connections when the detector gas flow
and/or the gas supply is turned on. The escaping gas can pose a health
and safety risk.
Turn on the gas supply and detector gas flow only when the nebulizer is
installed, the components in the gas flow path are properly connected,
and the protective cover of the gas filter assembly is installed.
NOTICE—Detector Gas Flow and Pump Flow
Improper handling of detector gas flow and pump flow start and stop
can damage the detector. Observe the respective order of steps:
•
When turning on detector gas flow and pump flow
Turn on the detector gas flow first, wait at least 5 minutes and only
then turn on the pump flow.
•
When turning off pump flow and detector gas flow
Turn off the pump flow first, wait at least 5 minutes and only then
turn off the detector gas flow. During the waiting period, the drain
pump drains any remaining mobile phase to waste before it
automatically turns off.
Preparations
1. Make sure that you observe the site requirements for the gas
supply. See
page 52
2. Turn on the gas supply to the detector. Observe the
Gas
specifications
for the gas supply pressure. See
section 9.1
Performance Specifications, page 164
Follow these steps
1. Turn on the gas flow in the detector using the
Gas
button on the
keypad of the detector or using Chromeleon.
2. Close the detector doors.
Summary of Contents for Vanquish VF-D20
Page 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Page 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Page 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Page 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Page 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Page 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Page 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...