9
•
Specifications
Page 164
Vanquish Charged Aerosol Detectors Operating Manual
9.1
Performance Specifications
The detector performance is specified as follows:
Type
Specification
Detection type
Charged aerosol detection
Nebulization
FocusJet, concentric
Pump flow rate
VH-D20
0.01 – 2.0 mL/min
VF-D20
0.2 – 2.0 mL/min
Evaporation temperature
VH-D20
Adjustable range: a5 to +100 °C
VF-D20
Selectable temperatures: +35 °C, +50 °C, +70 °C
Signals
Data collection rate
VH-D20
Adjustable, 2.0 Hz – 200 Hz (max. 100 Hz with Chromeleon 6.8)
VF-D20
Adjustable, 2.0 Hz – 100 Hz
Filter response time
Digital filter, selectable in numerical sequence (1-2-5 increments)
Digital filter constants
Selectable between 0.1, 0.2, 0.5, 1.0, 2.0, 3.6, 5.0 10.0 seconds
Dynamic range
Up to 4 orders of magnitude
Gas specifications
Gas pressure control
Electronically controlled
Supply gas
Compressed air or nitrogen
Gas supply pressure
482 - 551 kPa (4.8 - 5.5 bar, 70 - 80 psi)
Communication
USB
1 USB port (USB 2.0, "B" type connector)
1 USB hub with 3 ports (USB 2.0, "A" type connectors)
System Interlink
2 system interlink ports (RJ45-8 connectors)
Analog output
Installable option; 1 BNC-type analog output connector
Full-scale analog output
range
1 pA to 500 pA in 1-2-5 sequence
Analog signal output
0 – 1 V DC
Summary of Contents for Vanquish VF-D20
Page 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Page 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Page 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Page 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Page 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Page 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Page 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...