6
•
Operation
Page 114
Vanquish Charged Aerosol Detectors Operating Manual
Restarting the Detector
To restart the detector, follow these steps:
1. Prepare and restart the other modules in the Vanquish system,
following the instructions in the
Operating Manuals
for the modules.
Pay special attention to the
Preparing the Module for Operation
section.
2. Flush the components in the flow path before the detector before
you connect the detector to the system flow path.
3. Turn on the detector. Observe the following:
♦
If the detector remained in the system stack and all system
modules were turned off
Turn on the system with the system power button on the system
base. See
♦
If the detector is restarted after transport
Turn on the detector with the main power switch. See
4. Before starting an analysis, let the detector equilibrate and be sure
that it is ready for operation (see
section 6.5 Preparing the Detector
for Operation, page 92