6
•
Operation
Page 98
Vanquish Charged Aerosol Detectors Operating Manual
6.5.2
Verifying the Nebulizer Gas Pressure
Before operating the detector, verify the nebulizer gas pressure setting.
NOTICE
Operating the nebulizer with a gas pressure other than the
pressure specified for this nebulizer can damage the nebulizer and the
detector. After installation of the nebulizer, set the nebulizer gas
pressure to the pressure that is specified for your nebulizer. The
nebulizer gas pressure is defined on the gas pressure certificate that is
shipped with the nebulizer.
1. In Chromeleon, open the
Command
window. Navigate to the
Gas
Regulator Pressure Nominal
parameter.
2. Check if the gas pressure on the nebulizer certificate matches the
nebulizer gas pressure set in the
Gas Regulator Pressure Nominal
parameter.
If the nebulizer gas pressure deviates from the pressure on the
nebulizer certificate, enter the gas pressure from the nebulizer
certificate.
3. Close the
Command
window.
4. Monitor the
Gas Regulator Pressure
until it reaches the setpoint.
Summary of Contents for Vanquish VF-D20
Page 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Page 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Page 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Page 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Page 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Page 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Page 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...