Calibration
Defining the Response Factor
Thermo Fisher Scientific
TVA2020 Instruction Manual
4-13
Enter name:
RF3: Hexa
_________
Press char key
Use the alphanumeric keypad to enter a compound name for the response
factor. Press enter when you are finished.
If you choose 2=Numbers, the following display will appear:
FID
PID/FID
Enter RF3: Enter RF3:
FID:01.50 P&F:01.50
Enter=Accept Enter=Accept
Use the keypad to enter a new response factor and press ENTER to store it
in instrument memory.
Note
The TVA2020 will not accept a response factor of 00.00.
▲
Note
If a response curve is used, the TVA2020 will prompt you to enter
two numbers.
▲
FID
PID/FID
RF3: PROPANE RF3: PROPANE
FID: 1.25 1.50 PID: 1.25 1.50
FID: 1.25 1.50
1=Name 2=Numbers 1=Name 2=Numbers
Press 2=Numbers