Servicing
Replacing the Connector Access Door
7-26
TVA2020 Instruction Manual
Thermo Fisher Scientific
Use the following procedure to replace the connector access door (
Figure
7–22
).
1.
Cut off the retaining tail from the door if needed.
2.
Tilt the case top half up so that the cut off piece slips into the gap
between the board and the case.
3.
Take the new door and press the retaining tail into the hole. Use a
small blade screwdriver to work the tail into the hole.
Figure 7–22.
Replacing the Connector Access Door
Replacing the
Connector
Access Door