FID
PID/FID
FID: 24950ppm PID: 1.1ppm
FID: 2.49%
A--------------- A---------------
0=1 1=A 2=B 3=C 0=1 1=A 2=B 3=C
Once the tag information is entered, press the ENTER key. The screen
then changes to the next display. The instrument is now ready to start
logging. The screen displays live readings and the tag name. However, no
logging takes place until you enter the logging mode from the enhanced
probe or press the ENTER key on the instrument.
FID
PID/FID
FID: 24950ppm PID: 1.1ppm
FID: 2.49%
BLD01SITE5 BLD01SITE5
Enter=Start log Enter=Start log
Press ENTER or initiate the logging mode from the enhanced probe to
start the automatic logging. The instrument counts down from the time
selected in the SETUP/Log menu and stores the reading at the end of the
count down. For instance, if 12 seconds was selected as the logging time,
the instrument will count down from 12 and display
11,10,9,8,7,6,5,4,3,2,1,STORE. The instrument logs into memory either
the highest, average, or last reading during the countdown, as configured in
the SETUP/Log menu.
FID
PID/FID
FID: 24950ppm PID: 1.1ppm
FID: 2.49%
Sampling: 3sec Sampling: 3sec
Exit=Cancel Exit=Cancel
This action continues until the instrument shuts down due to low battery,
or the EXIT key is pressed, or the log memory is full.
Whenever the TVA2020 is in the RUN mode with Log VOC as the
logging selection, the instrument operates as a manual survey tool that logs
its readings into memory whenever you initiate logging. When the 1=Run
Run: Log VOC
Summary of Contents for TVA2020
Page 1: ...TVA2020 Instruction Manual Toxic Vapor Analyzer Part Number 111755 00 29Jul2014...
Page 4: ...Thermo Fisher Scientific WEEE Compliance...
Page 16: ...Tables xii TVA2020 Instruction Manual Thermo Fisher Scientific...
Page 76: ......
Page 96: ......
Page 118: ......
Page 190: ...Response Factors Response Factors C 4 TVA2020 Instruction Manual Thermo Fisher Scientific...