Optional Accessories
Enhanced Probe Option
8-6
TVA2020 Instruction Manual
Thermo Fisher Scientific
PID
0.66
ppm
Tag:
0.59ppm PID
2.50ppm FID
Sampling: 3s [Exit]
5.
Once EXIT is selected, you may enter another tag and begin logging
again or exit from the RUN mode.
Use the following procedure to use the enhanced probe in VOC logging
mode.
1.
In the RUN mode, from the Opening menu, press the
select
key to
“Enter logging mode.”
2.
Use the sidepack to enter a tag, or simply select “Accept” to enter a
blank tag.
3.
Choose “Start log sampling” to begin Auto logging.
4.
The Enhanced probe will display the count down for the interval
selected in the SETUP menu and freeze on the reading to be saved
(Maximum, Average, or Last Reading).
PID
0.66
ppm
Tag:
0.59ppm PID
2.50ppm FID
[Sav]
–
Again -Exit
5.
You may choose to save the reading (SAV), repeat the reading for the
same tag (AGAIN), or return to the previous menu (EXIT).
VOC Logging with
the Enhanced Probe