Calibration
Calibration Configuration
Thermo Fisher Scientific
TVA2020 Instruction Manual
4-5
Note
If your instrument is equipped with dual detectors, you may choose
to calibrate the FID separately or PID/FID together.
▲
Note
If your instrument is configured for multiple span points, be sure to
set the concentration for ALL span points.
▲
1.
From the CAL CONFIG MENU (Page 1) display, press 2=Span concs.
The upper display (or two displays if the unit is a dual detector version)
will display the concentration value of your span gas (expressed as ppm
or %) as of the last calibration:
FID
PID/FID
FID: 500ppm PID: 100ppm
FID: 500ppm
Span Gas Concent Span Conc:1=Both
Enter=New conc 2=PID 3=FID
If the TVA2020 is configured for multiple calibration points, the span gas
concentration values for Point #3 will be displayed. The Up and Down
arrow keys can be used to scroll through the span gas concentration values
for other points:
FID
PID/FID
FID: 500ppm PID: 100ppm
FID: 500ppm
Span Pt3 (Up/Dn) Span Pt3 (Up/dn)
Enter=New conc 1=Both 2Pid 3Fid
2.
To change a span gas concentration value, choose the detector(s) of
interest if using a dual detector analyzer, or press ENTER if using a
single detector analyzer. The following display will appear:
Summary of Contents for TVA2020
Page 1: ...TVA2020 Instruction Manual Toxic Vapor Analyzer Part Number 111755 00 29Jul2014...
Page 4: ...Thermo Fisher Scientific WEEE Compliance...
Page 16: ...Tables xii TVA2020 Instruction Manual Thermo Fisher Scientific...
Page 76: ......
Page 96: ......
Page 118: ......
Page 190: ...Response Factors Response Factors C 4 TVA2020 Instruction Manual Thermo Fisher Scientific...