Troubleshooting
Troubleshooting Guide
Thermo Fisher Scientific
TVA2020 Instruction Manual
6-3
Problem
Possible Reason
Solution/Action
Broken ignitor
Inspect ignitor coil on FID
capsule for breakage. If
broken, replace capsule.
Dirty ignitor contacts
Inspect/clean ignitor contacts
(
not
ignitor coils).
Battery charge low
Charge battery pack.
FID noisy
Water/contamination in the
detector chamber
Clean/replace FID capsule
and flame detector.
Erratic pump flow
Clean/replace cups.
Clean/replace flame arrestor.
Service
pump.
Bad calibration
Ensure proper calibration.
Unable to calibrate FID
FID flame out
Ignite FID.
Span concentration not
properly set
Input correct span gas
concentration at CAL menu.
Cal gases contaminated
Use clean cal gases and
sampling equipment.
Sample line/filter cups
contaminated
Clean/replace sample line.
Clean/replace filter cups.
Service
pump.
FID capsule contaminated or
faulty
Clean/replace FID capsule.
Contaminated hydrogen tank
Replace hydrogen tank.
Internal detector fault or
contamination
Call Thermo Fisher Scientific
Service Department.
Excessive hydrogen
consumption (<8 hours of
run time for 2200 psi
hydrogen)
Insufficient hydrogen pressure
Refill tank.
Leaking hydrogen tank
Replace tank.
Internal hydrogen leak
Call Thermo Fisher Scientific
Service Department.
Flameout problems
Sample hydrogen content too
high
Use dilutor kit to achieve
concentration within the
dynamic range.
Insufficient oxygen in the
sample (<14%)
Use dilutor kit to dilute
sample with air
contaminating sufficient
oxygen.
Summary of Contents for TVA2020
Page 1: ...TVA2020 Instruction Manual Toxic Vapor Analyzer Part Number 111755 00 29Jul2014...
Page 4: ...Thermo Fisher Scientific WEEE Compliance...
Page 16: ...Tables xii TVA2020 Instruction Manual Thermo Fisher Scientific...
Page 76: ......
Page 96: ......
Page 118: ......
Page 190: ...Response Factors Response Factors C 4 TVA2020 Instruction Manual Thermo Fisher Scientific...