Servicing
Cleaning the PID Lamp
Thermo Fisher Scientific
TVA2020 Instruction Manual
7-9
Use the following procedure to remove and clean the PID lamp. Replace
when needed.
WARNING
Power down the instrument before performing any service
procedures.
▲
WARNING
To prevent a potential explosion, do not operate the
instrument with the PID compartment open or FID detector cap
removed.
▲
WARNING
All maintenance procedures must be performed on a clean
surface using clean tools. Avoid touching the lamp's window as well as the
metalized portion of the cell assembly with your bare fingers. Fingerprints
left on these parts may adversely affect the sensors operation. Latex gloves
are preferred, but if they are not used, your hands must be clean and free of
oils, lotions, etc. It is acceptable to hold the lamp by its glass body or by the
edges of the window.
▲
1.
Turn the instrument OFF and remove the sensor from the instrument
(
Figure 7–3
).
Figure 7–3.
Removing the PID Sensor
2.
Remove the filter cap by applying slight upward pressure with the tip of
a screwdriver or X-Acto blade just below the hole in the cap and
between the cap and housing; it will pop off (
Figure 7–4
).
Cleaning the PID
Lamp
Summary of Contents for TVA2020
Page 1: ...TVA2020 Instruction Manual Toxic Vapor Analyzer Part Number 111755 00 29Jul2014...
Page 4: ...Thermo Fisher Scientific WEEE Compliance...
Page 16: ...Tables xii TVA2020 Instruction Manual Thermo Fisher Scientific...
Page 76: ......
Page 96: ......
Page 118: ......
Page 190: ...Response Factors Response Factors C 4 TVA2020 Instruction Manual Thermo Fisher Scientific...