Servicing
Cleaning the PID Lamp
Thermo Fisher Scientific
TVA2020 Instruction Manual
7-11
Figure 7–7.
Removing the Cell Assembly
6.
With fine tipped tweezers, grasp the lamp by placing the tips in the
housing notch and gently pull it out. Be careful not to scratch the lamp
lens or chip edges (
Figure 7–8
).
Figure 7–8.
Removing the PID Lamp
7.
Grab the lamp by the cylindrical glass body and clean the window by
rubbing it against the polishing pad. Use a circular motion and try to
keep the window surface flat relative to the pad. Five seconds of
rubbing will be enough in most cases. Another indication of cleaning
completeness is that you have used about 1/6-inch of the pads surface
during the procedure (
Figure 7–9
).
Figure 7–9.
Cleaning the PID Lamp