Optional Accessories
Enhanced Probe Option
8-10
TVA2020 Instruction Manual
Thermo Fisher Scientific
If you choose a leak source and repair menu, be sure to select “Sav” when
you return to this menu so the reading and repair information will be saved
in memory before proceeding to the next component record.
Displays containing additional information about the component and the
TVA2020 status are available from the handheld unit. If you press and
hold the MENU key for approximately 2 seconds, the Enhanced display
will present an INFO page:
PID
0.66
ppm
Tag:0020
Area:1515 BEN
Subarea:WEST
Leak: 10
RF:1.00
14:17:52 [Pg] -Exit
The INFO page overwrites the bottom four lines of the record with other
fields such as Location (Loc) and Description (the two lines below the
location). The Location field contains the Area and the Subarea where the
component is located. The Description field displays up to 40 characters of
additional information about the component. Selecting “Pg” at the bottom
of the display will step you through other INFO pages containing
information such as Run ID, Record number, number of points logged,
number of leakers found, number of repairs performed, pump status, free
memory, etc. Selecting “Exit” will return you back to the normal
component record.
Additional Component
Information
Summary of Contents for TVA2020
Page 1: ...TVA2020 Instruction Manual Toxic Vapor Analyzer Part Number 111755 00 29Jul2014...
Page 4: ...Thermo Fisher Scientific WEEE Compliance...
Page 16: ...Tables xii TVA2020 Instruction Manual Thermo Fisher Scientific...
Page 76: ......
Page 96: ......
Page 118: ......
Page 190: ...Response Factors Response Factors C 4 TVA2020 Instruction Manual Thermo Fisher Scientific...