Servicing
Cleaning the PID Lamp
7-10
TVA2020 Instruction Manual
Thermo Fisher Scientific
Figure 7–4.
Removing the Filter Cap
3.
With fine-tipped tweezers, remove both filter media and set aside
(
Figure 7–5
).
Figure 7–5.
Removing the Filter Media
4.
Using the X-Acto blade, remove the spacer and set aside (
Figure 7–6
).
Figure 7–6.
Removing the Spacer
5.
With fine tipped tweezers, carefully remove the cell assembly by prying
under the cell’s edge where connector pins are located (
Figure 7–7
).