Servicing
Servicing the Pump
Thermo Fisher Scientific
TVA2020 Instruction Manual
7-19
Servicing the pump involves removing the pump assembly, accessing the
orifice, and replacing the pump.
Use the following procedure to remove the pump assembly.
1.
Disconnect the 8 pin connector from the pump board (
Figure 7–17
).
2.
Using the correct wrenches, remove the hex elbow attached to the valve
assembly (
Figure 7–17
).
3.
Remove the Tygon tubing from the barb fitting on the sample inlet
assembly (
Figure 7–17
).
Figure 7–17.
Removing the Pump
4.
Lift the pump up and out of the instrument.
5.
Disconnect the pump motor from the pump board.
Servicing the
Pump
Removing the Pump
Assembly
Tygon Tubing
8 pin Connector
Hex Elbow