A-5
•
In the “one’s” place of the status code, a status code of “2” is displayed.
In the “one’s” place of the status code table, the “2” status code matches
the (H)2 “Power Switch (TPIC)” status code. This is one of the status
codes that the unit is displaying in its status code.
•
In the “ten’s” place of the status code, there are no (0) status codes
displayed.
•
In the “100’s” place of the status code, there are no (0) status codes
displayed.
•
In the “1,000’s” place of the status code, a status code of “B” is displayed.
Because there are no status codes in the status code table that match this
number, the numbers needs to be broken down further. Convert “B” to a
decimal number using Table A-2 — which converts B to “11.” Next, look
at the status code table to decipher the status code. In the “1,000’s” place
of the table, there are three status codes that, when added together, will
total 11: (H)1000 “Serial Port Problem,” (H)2000 “Line Printer Problem”
and (H)8000 “Leak on Loop B.” These are three more of the status codes
that the unit is displaying in its status code.
•
In the “10,000’s” place of the status code, there are no (0) status codes
displayed.
•
In the “100,000’s” place of the status code (800000), a status code of
“800000” is displayed, which matches: (H)800000: “Sensor Comm Prob-
lem.”
•
In the “1,000,000’s” place of the status code, there are no (0) status codes
displayed.
Therefore, the downloaded status code “8433666” ((H)80B002) breaks
down into the following status codes, according to the example status code
table:
(H)2 “Power Switch (TPIC)”
(H)1000 “Serial Port Problem”
(H)2000 “Line Printer Problem”
(H)8000 “Leak on Loop B”
(H)800000: “Sensor Comm Problem.”