5. Place a piece of plastic or another protective material over the exposed
TEOM filters.
6. Using a soapy water, alcohol or freon solution, clean the left side of the
air inlet (Figure 5-31). A soft brush may be used to remove particulate
matter on the insides of the walls.
7. Allow the air inlet to dry.
8. Remove the protective material from the exposed TEOM filters.
9. Close the mass transducer and latch the latch.
10. Install the air thermistors into the cap of the mass transducer assembly
and tighten lightly with the wrench.
11. Close and latch the door to the unit. Keep the door open for as short a
time as possible to minimize the temperature change in the system.
12. Turn on the TEOM 1405 unit.
Figure 5-31.
Figure 5-31.
Figure 5-31.
Figure 5-31.
Figure 5-31.
Air inlets.
Nozzle