7. The Remove Sample Filter screen will display. Open the mass
transducer and remove the standard TEOM filter from the mass
transducer. (Refer to the section on installing/removing TEOM filter
for assistance on removing the filter.) Close the mass transducer
(without installing another filter). Select the
Next >
button.
Note.
Note.
Note.
Note.
Note. DO NOT use the calibration filter exchange tool for
installing or removing ANY filter other than the pre-weighed
calibration filter.
▲
8. The Stabilizing screen will display (Figure 5-65). While the instrument
waits to mesaure the frequency of the system with no TEOM filter
installed, a countdown timer on the screen will show the progress of
the stabilization step. When the stabilization “Complete” message
displays, select the
Next >
button.
Figure 5-65.
Figure 5-65.
Figure 5-65.
Figure 5-65.
Figure 5-65.
Stabilizing screen.
Summary of Contents for TEOM 1405