3. When in the Data Storage screen, select the Edit List button to display
the Edit Data Storage screen. Press the names of the variables you wish
to log, up to a maximum of 20. Use the
Next Page >
and
< Previous
Page
buttons to scroll through the entire list of variables which can be
stored. Select the OK button when all the desired variables have been
selected.
4. Use the
∨∨∨∨∨
and
∧∧∧∧∧
buttons to scroll through the list of selected variables
to ensure that all desired variables are selected.
5. When in the Data Storage screen, select the
Storage Interval
button to
select the interval for data storage. Enter the desired data storage
interval into the keypad and select the
Enter
button. For example, if
the storage interval is 10 seconds, every 10 seconds the instrument will
log (save) the data in the selected variables.
6. When all the desired variables are selected and the Storage Interval is
set, select the
< Back
button to return to the Settings screen.
2. When in the Settings screen, select the
Data Storage
button to display
the Data Storage screen (Figure 3-10).
Figure 3-10.
Figure 3-10.
Figure 3-10.
Figure 3-10.
Figure 3-10.
Data Storage screen.