11. The Seat Filter screen will display. Gently press down on the TEOM
filter to ensure that it is seated properly (Figure 5-9).
12. Remove the filter exchange tool by slowly retracting it until it clears the
filter (Figure 5-9). Do not disturb the filter.
13. Place the bottom of the filter exchange tool on top of the TEOM filter
(Figures 5-9) and apply downward pressure (approximately 0.5 kg or 1
lb) to seat the filter firmly in place.
14. The Precondition Filter screen will display. Place a new TEOM filter
on the empty conditioning post on the mass transducer. (Refer to the
next section for more information on conditioning filters.) Select the
Next >
button.
15. The Close Instrument screen will display. Raise the mass transducer to
the closed position and fasten the holding rod onto the latch plate.
16. Close and latch the door to the sensor unit. Keep the door open for as
short a time as possible to minimize the temperature change in the
system. Select the
Next >
button.
Figure 5-9.
Figure 5-9.
Figure 5-9.
Figure 5-9.
Figure 5-9.
Placing the filter on the tapered
element.