, TEOM 1405
5-23
7. Unscrew the filter mount for the bypass flow channel.
8. Slide the large filter cartridge off the mount and install a new cartridge
onto the mount (Figure 5-29).
9. Install the mount into the unit, then install the cover for the bypass
flow.
10. Plug in the sample pump and return to normal operation.
Clean filter cartridge
Figure 5-29.
Figure 5-29.
Figure 5-29.
Figure 5-29.
Figure 5-29.
Bypass filter mount with clean
and dirty filter cartridges.
Dirty filter cartridge