7. The Remove Inlet screen will display. Remove the inlet (Figure 3-4).
Select the
Next >
button.
8. The Attach Audit Adapter screen will display. Attach the leak check/
flow audit adapter to the top of the flow splitter.
9. Slowly close the valve on the leak check adapter. Select the
Next >
button.
10. The Stabilizing screen will display. Allow 1 minute for the flows to
stabilize then select the
Next >
button.
11. The Replace Inlet screen will display. Slowly open the leak check valve
to restore flow to the system. Remove the flow audit/leak check
adapter and attach the inlet to the top of the sample inlet tube (Figure
3-4). Select the
Next >
button.
Inlet
Flow splitter
Figure 3-5.
Figure 3-5.
Figure 3-5.
Figure 3-5.
Figure 3-5.
Leak check adapter.
Leak check
adapter
Valve closed
Figure 3-4.
Figure 3-4.
Figure 3-4.
Figure 3-4.
Figure 3-4.
Inlet assembly.