RadEye GX, GX-L
DB-074 E
8-13
Pi/Ff 14.09.2016
8.5.5
Probe configuration
Figure 8-9: Probe information
Probe type:
This field can be edited between customized and
factory set probes. For factory set probes it is only possible to
change the calibration factor. For user defined individual detec-
tors every field can be edited
Name:
this column is used to set a name for the probe. This name
is displayed after start up the RadEye and in submenu “Select
counter tube”. It is highly recommended to include the serial
number or other unique identification label.
Cal.Factor:
Calibration factor for dose rate. If the calibration
factor is not known, use calibration factor “0”. In this case “###”
is displayed in dose rate mode (see chapter 5.3).
Dead time:
In these fields a dead time correction factor for each
counter can be edited.
Dead time coefficient:
Quadratic correction in ns².