RadEye GX, GX-L
DB-074 E
15-1
Pi/Ff 14.09.2016
15. Technical data
15.1
RadEye GX
Measured
quantities:
Count rate (cps, cpm)
Surface contamination (Bq, dps, dpm, Bq/cm
2
)
Dose rate (R/h, Sv/h, rem/h, Gy/h)
Measuring range:
Default: 10,000 cps
Largest indication
Count rate: 9999 kcps, 9999 kcpm,
Activity: 9999 kBq, 9999 kdps, 9999 kdpm,
9999
kBq/cm²
Dose rate: 9999 R/h, 9999 rem/h, 9999 Sv/h
Probe cables
RG 58, Max. 2 m – MHV connector,
High voltage range
350V…1100V
Output impedance
1M
Ω
Maximum number
of counter tube
16 counter tubes with adjustable high voltage,
calibration factor, dead time correction, overload
threshold, detector area and timeout for detector
failure. Default values for HP-190A, HP-210,
HP-270, HP-290,HP-360 and EP-15 are available
Alarm threshold:
Two alarm thresholds for count rate, activity,
dose and dose rate each.
Default setting: see chapter 2.4
Audible alarm
intensity:
At least 80 dB(A) at a distance of 30 cm