Pi/Ff 14.09.2016
Open next the serial interface step window in the RadEye.EXE
software:
Configuration / Com settings...
menu. The used serial
port has to be selected next.
10.6
RadEye lost date and time settings
The date and time information is lost, each time the batteries are
removed or below the required voltage for operation. The sim-
plest way to program the date and time is to use the “synchro-
nize” function of the optional RadEye.EXE software. Please refer
to chapter 8.5.2. Else the current date and time can be set also
through the RadEye’s menu “Settings” / “Set Date / Time”
10.7
RadEye’s battery status is lower than expected
Please check if the correct battery type is selected under “set-
tings” / “Batt. type” and change the type in case. Rechargeable
batteries supply a voltage of 1.2V each, while Alkaline batteries
provide 1.5V.
Rechargeable batteries offer also a more steep drop from accepta-
ble battery voltage to “battery low” than Alkaline types.
10.8
History data shows the wrong time and date
See chapter 8.5.2.