RadEye GX, GX-L
DB-074 E
8-15
Pi/Ff 14.09.2016
correction only. For individual calibration and linearity check of
dose rate detectors, the PC-software CalRadEye.EXE should be
used. Please note that the cable type and length have an influence
on the calibration parameters.
Typical parameters
Detector
HP-270 HP-210 HP-
190A
HP-360 EP-15
HV
[V] 950 900 900 900 500
cal factor
[µR/h/cps]
50 20 31 20 19
DT
[µs] 95 50 170 50 70
QC [ns]
-3.1
-0.6 -9.5 -0.6 0
Over range
[cps] *)
50,000 30,000 20,000 30,000 30,000
Over range
power
180 2,000
250 2000 300
Area
[cm²]
0 20 7 20 20
Timeout
[s]
100 60 60 60 60
*) Over range [cps] for RadEye GX-L is limited to 10,000 cps.