RadEye GX, GX-L
DB-074 E
0-9
Pi/Ff 14.09.2016
SAFETY INSTRUCTIONS
Dose integration is performed only, if the instrument is in a
dose rate measuring mode (R/h, Rem/h, Sv/h, Gy/h). In case
the measuring range is exceeded, the value of the upper end of
the dose rate measuring range is used for the dose integration.
The measuring value and range depends on the properties of
the connected external detector. It is the responsibility of the
user or supervisor to select the correct probe.
Extension of the probe specific default measuring range can be
achieved by individual calibration and determination of the
dead time behavior.
The RadEye GX, respectively the connected Geiger Mueller
probes are not well suited for pulsed radiation. Depending on
the sensitivity of the detector and the conditions of the pulsed
field, there can be a significant under-response.
Do not use the unit if error messages appear on the screen.
The earphone connector at the bottom side of the instrument
must be exclusively used by equipment that is specified for
use with RadEye GX.