15-2
DB-074 E
RadEye GX, GX-L
Pi/Ff 14.09.2016
Working
temperature:
-20°C ... + 50°C
Storage
temperature:
-25°C … + 50°C
Ambient air
pressure:
700 hPa … 1100 hPa
Relative humidity:
10 ... 90 % at 35°C not condensing
Operating voltage:
1,8 ... 4 V,
Battery low voltage starting from 2.3 V
Degree of
protection:
IP 32 according to EN 60 529
EMC:
Disturbance emission : EN 61000-6-3
Immunity : EN 61000-6-2
Size:
110 mm x 67 mm x 62 mm
With rubber protection, without cable
Weight:
around 150g including AAA cells and pro-
tection sleeve
Internal memory:
The last 1500 measured values are saved and
can be read out via PC program. Max- and
mean value of count rate and dose rate. The
time interval is factory preset to 120s by
default.
Logbook with 250 entries for changes of
configuration, occurring alarms and errors.
Averaging filters:
Ratemeter ADF (
A
dvanced
D
igital
F
ilter)
Digital RC-Filter with time constant
1s.....180s, depending on dose rate and dose
rate changing.
Ratemeter Tau: Linear Ratemeter with ad-
justable time constant from 1s to 60s