RadEye GX, GX-L
DB-074 E
15-5
Pi/Ff 14.09.2016
Internal memory:
The last 1500 measured values are saved and
can be read out via PC program. Max- and
mean value of count rate and dose rate. The
time interval is factory preset to 120s by
default.
Logbook with 250 entries for changes of
configuration, occurring alarms and errors.
Averaging filters:
Ratemeter ADF (
A
dvanced
D
igital
F
ilter)
Digital RC-Filter with time constant
1s.....180s, depending on dose rate and dose
rate changing.
Ratemeter Tau: Linear Ratemeter with ad-
justable time constant from 1s to 60s
Power consump-
tion:
1.4 mA: normal operation without alarm
signals and LCD illumination. High voltage:
900V
27 mA with illuminated LC display
9 mA LED alarm
33 mA acoustic alarm
26 mA vibrator alarm
Battery service
life:
500h using two alkaline AAA cells de-
pending on the operating mode
250h using 800mAh NiMH accumulator