3-16
DB-074 E
RadEye GX, GX-L
Pi/Ff 14.09.2016
3.3.2
Accumulated counts
With accumulated counts selected via radeye.exe the RadEye
show the collected counts within the measurement time in place
of the mean value. This option is only available in combination
with measurement unit “cps”
3.4
Background measurement
To enter background measurement select in submenu “Back-
ground” “Start Bgr.meas”.
The background measurement starts with the parameter set in
submenu option „Set parameter“. To start the measurement press
the up key. While the measurement is in progress “Exit” is blink-
ing.
The measurement is finished if either the elapsed time arrived the
preset time or the accumulated counts reached preset count.
Action of
“Info” key
Measure-
ment value
Elapsed time
Preset time
Accumulated
counts
Preset count