Thermo Scientific Orion Star LogR™ Meter User Guide
Autosampler Interfacing
If an ORP electrode will be calibrated on the autosampler, turn
the ORP calibration feature on by selecting YES. The meter must
perform an automatic calibration with the autosampler, so only
Thermo Scientific Orion ORP standard can be used. This menu will
be shown if the number of pH calibration points is set to zero.
If an ISE electrode will be calibrated on the autosampler, enter the
number of ISE calibration points. Zero, two or three points can
be selected. Set the number of calibration points to zero if an ISE
calibration with the autosampler is not needed. Once the number of
ISE calibration points is selected, enter the calibration standard values
in the setup menu. This menu will only be shown if the meter has ISE
function and the number of pH calibration points is set to zero.
Press
twice to move the arrow icon to the middle line
and press
to display the first standard value. Press
to
move the arrow icon to the top line and enter the value of the
first standard. Press
until the first digit to be changed is
flashing, press
/
to change the value of the flashing digit
and continue to change the digits until the meter displays the
concentration value of the standard. Once the standard value is set,
press
until the decimal point is in the correct location. Repeat
this procedure for the second and third standard (if used).
Press
twice to move the arrow icon to the middle line and then
press
to display the next setup parameter.
Enter the total number of sample beakers to be measured. Do not
include rinse beakers, calibration beakers or empty beakers.
Note:
The error 005 message indicates that a bad value was entered. Press
to return to
the setup menu and check the allowed value range.