29
Thermo Scientific Orion Star LogR™ Meter User Guide
LogR and Automatic Temperature Compensation (ATC) Calibration
ATC Probe Calibration Technique
The temperature calibration mode of the Orion Star LogR meter allows the
temperature measurement of the ATC probe to be manually adjusted.
1. Connect an ATC probe to the meter and change the temperature input to ATC in
the setup menu.
2. In the measurement mode, press
to choose the top measurement line and
press
/
until the temperature is shown for the selected line.
3. Press
to begin the calibration.
4. When the reading stabilizes, the arrow icon and the first digit will flash. Enter
the temperature by pressing
/
to adjust each digit and
to move to
the next digit.
5. Press
to save and end the calibration.