69
Thermo Scientific Orion Star LogR™ Meter User Guide
Autosampler Interfacing
Meter Preparation for Operating the Autosampler
1. In the measurement mode, change each display line to the desired parameter
and blank any unneeded lines. See the Selecting the Measurement Parameter
section for detailed instructions.
2. Review and change the parameters for basic meter functions in the setup menu.
Note:
The first time the meter is prepared for use with the autosampler, confirm all of the
applicable setup menus. As long as the meter is connected to a power supply, the setup
parameters do not need to be reprogrammed. To prevent the meter from losing the programmed
parameters in the event of a power failure, install four AA alkaline batteries in the meter.
If a pH electrode will be calibrated on the autosampler, select the
USA or EUrO buffer set that will be used for automatic calibration.
If an ion selective electrode (ISE) will be calibrated on the
autosampler, select the ISE units that will be used for calibration and
analysis.
Set the read type on the meter to AUTO-READ.
Turn the automatic datalogging feature On or OFF.
Turn the datalog deletion setting to nO or YES. See the
Datalog
and Calibration Log
section for details on setting this parameter.
If a stirrer probe will be used, set the stirrer speed. The stirrer speed
can be set from one to seven or OFF.