37
Thermo Scientific Orion Star LogR™ Meter User Guide
ISE Technique
ISE Temperature Display and Calibration
ISE Temperature Display
Orion Star LogR meters allow the temperature to be viewed on individual
measurement lines in addition to the temperature display on the top, left of the
screen.
To view the temperature for the ISE measurement line:
1. Press and hold
until the meter displays the measurement mode.
2. Press
to choose the top display line. The arrow icon will point to the
selected line.
3. Press
/
to change the value on the selected line. The top line can
be changed to display pH (pH), millivolts (mV), relative millivolts (RmV),
concentration (ISE), temperature (no icon) or a blank line.
ATC Probe Calibration Technique
The temperature calibration mode of the Orion Star LogR meter allows the
temperature input for the ATC probe to be manually adjusted.
1. Connect an ATC probe to the meter and change the temperature reading and
compensation setting to ATC in the setup menu.
2. In the measurement mode, press
to choose the top measurement line and
press
/
until the temperature is shown for the selected line.
3. Press
to begin the calibration.
4. When the reading stabilizes, the arrow icon and the first digit will flash. Enter
the temperature by pressing
/
to adjust each digit and
to move to
the next digit.
5. Press
to save and end the calibration.