Analyzer Operation
III-27
Thermo Scientific Orion 2120XP Calcium Hardness Analyzer User Guide
Default
Operator Action
Scrolling Marquee
Main Display
Notes
mADJ
mADJ
(flashing)
SEL
SCrn
SEL CH1 or SEL
CH2 on main
display of two
channel analyzer
Press
to set the sensor and
temperature mA offset values
SELECT TEMP OR SENSOR
OUTPUT
SOUt or tOUt
(flashing)
Press
/
to scroll between
SOUt or tOUt and select SOUt
SELECT TEMP OR SENSOR
OUTPUT
SOUt
(flashing)
Press
to accept the setting and
advance to the next screen
0.00
SENSOR 4-20mA OUTPUT
ADJUSTMENT
AdJ
0 . 0 #
(last flashing digit)
Press
/
to set the sensor
mA offset value
Press
to set a positive or negative
sensor mA offset
SENSOR 4-20mA OUTPUT
ADJUSTMENT
AdJ
0 . 0
#
(change flashing digit)
Press
to accept the setting and
return to the main setup mode
CH1 or CH2
(flashing)
SEL
SCrn
Displayed for two
channel analyzer
Press
to return to the channel
specific menu options in the setup mode
Action required
for two channel
analyzer
mADJ
(flashing)
SEL
SCrn
SEL CH1 or SEL
CH2 on main
display of two
channel analyzer
Press
to set the sensor and
temperature mA offset values
SELECT TEMP OR SENSOR
OUTPUT
SOUt or tOUt
(flashing)
Press
/
to scroll between
SOUt or tOUt and select tOUt
SELECT TEMP OR SENSOR
OUTPUT
tOUt
(flashing)