The setup mode features programmable menu options. The order of the
menu options is dependent on the direction the operator scrolls by pressing
or
. The menu options are listed below by pressing
.
The following menu options are displayed in the main setup mode of one
and two channel analyzers.
Set either of two password options:
• System password – Setup settings protected, accessed by authorized
operators only
• Calibration password – Calibration menu data is protected, accessed by
authorized operators only
• Default password is 0000 – Disables both passwords
• Forgot your password? Contact Technical Support at 1-800-225-1480
Set the date in US or Europe format:
• Enter month, day and year
• Default date – System will continue to keep date and time due to
battery back up, operator must set in accordance to local time zone
• If the battery is removed, the system will show 01/01/2000
Set the time:
• Enter hour and minutes in 24 hour format
• Default time – System will continue to keep date and time due to
battery back up, operator must set in accordance to local time zone
• If the battery is removed the system will show 00:01
Set the data logging interval for measure log (calibration and error logs are
accessed in the test mode):
• Set the log interval as hour:minute
• Default log interval is 00:00 – logging disabled
• Minimum log interval is 1 minute, maximum log interval is 99 hours
and 59 minutes