Thermo Scientific
Dionex IC Pure system |
69
System control
|
Chapter 12
Press the “Menu“ key three times. The display shows:
Ultrapure cartridge serial number
Press the “Menu“ key four times to enter the serial cartridge number. The display shows:
Rinsing procedure
The fifth press of the "Menu" button enters the Rinse mode. This rinses or flushes the unit to drain.
During the rinse, the pump and solenoid valve (V4) are activated for a programmed period of time as
set in the OEM menu.
(see “Set the interval rinse time”
Note
You can find more information in chapter “Replacing the UV lamp” on page 89
Note
You can find more information in chapter “Changing the ultrapure cartridge” on page 79.
Step
Action
Figure
1
Press the “Menu“ key five times.
The display shows:
2
Confirm rinse by pressing the
“Enter“ button. The rinsing is started
for 30 sec.
Summary of Contents for dionex ic pure water
Page 10: ......
Page 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Page 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Page 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Page 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Page 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Page 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Page 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Page 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Page 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Page 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Page 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Page 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Page 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Page 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Page 127: ...Thermo Scientific IC Pure 117 ...