Technical specifications
14
|
Dionex IC Pure system
Thermo Scientific
Chapter 5
Product water quality
Conductivity (reference tempe25 °C)
0.055 µS/cm
Resistance
up to18.2 M
·cm
TOC value
max. 5 ppb (µg/L)
RNase
DNase
< 0.003 ng/mL
< 0.4 pg/uL
Bacteria
< 0.1 CFU/mL
Bacterial endotoxines
0.001 EU/mL
Particles > 0.2 µm per mL
< 1 µm/mL
Flow rate without pretreatment cartridge at manually
operated point of use (POU)
0.6 L/min ± 0.2 L/min
Flow rate with pretreatment cartridge at manually
operated POU
0.2 L/min + 0.2 L/min
Dimensions and weight of Dionex IC Pure system (50132810) with
ultrapure cartridge (09.1006)
Height
550 mm (21.65 in)
Width
306 mm (12.05 in)
Depth
385 mm (15.16 in)
Base area
277 x 372 mm (10.91 x 14.65 in)
Weight including ultrapure cartridge
26 kg (57.2 lb)
Weight with full tank
32 kg (70.55 lb)
Dimensions and weight of external 20 L tank with level control
(50146378)
Height
490 mm (19.29 in)
Width
190 mm (7.48 in)
Depth
485 mm (19.09 in)
Base area (height x width)
340 x 483 mm (13.39 x 7.48 in)
Weight of tank and filling station
7 kg (18.07 lb)
Weight with full tank
27 kg (60 lb)
Summary of Contents for dionex ic pure water
Page 10: ......
Page 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Page 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Page 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Page 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Page 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Page 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Page 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Page 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Page 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Page 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Page 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Page 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Page 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Page 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Page 127: ...Thermo Scientific IC Pure 117 ...