System control
68
|
Dionex IC Pure system
Thermo Scientific
Chapter 12
Set the limiting value using the “Arrow“ buttons (Nonstop button to increase the number, Interval
button to decrease and UV button to move to the right). With settings above 50 µS/cm (0.02
M
·cm), the limiting value is switched off and the word off appears in the display.
Press the “Menu“ key once. The display shows:
Ultra pure water limiting value
Press the “Menu“ key twice to set the ultrapure water limit value. This sets an error message to occur
when the system reaches the lowest acceptable level of purity. As soon as the error display is
switched on, the error will be displayed in both “Interval“ and “Nonstop“ modes. When the error
display is switched off, the error is only displayed in “Nonstop“ mode. The 'Lim. val.pure w.'
message is displayed when the limiting value is exceeded.
Ultrapure water measuring range: 0.1 M
·cm
Limiting value setting range:0.055- 5.0 µS/cm
Basic setting: 0.2 µS/cm (10 M
·cm)
Basic setting, error suppression: On
With settings above 5.0 µS/cm, the limiting value is switched off and the word 'Off' is shown in the
display.
Press the “Menu“ key two times. The display shows:
UV-Lamp operating time
In this menu, the operating hours of the UV-lamp are displayed. When the maximum operating time
is reached, the “UV-time“ error message is triggered. The determination of UV-intensity is not
possible for the Dionex IC Pure system. Therefore the display remains permanently as
“UV-intensity - - -“.
Summary of Contents for dionex ic pure water
Page 10: ......
Page 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Page 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Page 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Page 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Page 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Page 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Page 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Page 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Page 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Page 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Page 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Page 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Page 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Page 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Page 127: ...Thermo Scientific IC Pure 117 ...