Thermo Scientific
Dionex IC Pure system |
29
Installation
|
Chapter 7
Step
Action
Figure
1
a.
Remove the tube jumper from the
“IN“
connections
on the right side of the Dionex IC Pure system.
b.
Attach the “connecting hose 20 L tank to IC Pure“
supplied with the unit to the
“IN“
connection. Attach
the „“Drain hose IC Pure (d8 mm (31 in) with quick
disconnect)“ to
“OUT“
connection on the unit. Route
the tube from the
“OUT“
nozzle with a downward
slope to the drain
.
c.
If you are using a pretreatment cartridge in the unit,
remove the tube jumper from the left concentrate
connection on the Dionex IC Pure system and route a
o.d. 8 mm (0.31 in) tube to the left connection of the
20 L tank.
d.
Attach the tube from the 'IN' nozzle on the unit to one
of the two connections on the external 20 L tank.
e.
Attach the o.d. 8 mm (0.31 in) tube into the right
connection of the external 20 L tank for tank overflow
and route the tube with a downward slope to the
drain.
Note
If you are not using a pretreatment cartridge,
skip to step d.
Note
Either of the two connections on the external 20
L tank may be used to attach the tube to the unit.
IN / OUT Connections
o.d. 8 mm (0.31 in) tubes
a.)
b.)
d.)
20 L tank
connections
Concentrate
connection
c.)
20 L tank
connection for
tank overflow
e.)
o.d 8 mm (0.31 in) tube coming from
concentrate connection of Dionex IC Pure
system. Connected only if pretreatment
cartridge is used with IC Pure system.
o.d. 8 mm (0.31
in) tube
Summary of Contents for dionex ic pure water
Page 10: ......
Page 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Page 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Page 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Page 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Page 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Page 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Page 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Page 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Page 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Page 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Page 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Page 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Page 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Page 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Page 127: ...Thermo Scientific IC Pure 117 ...