43
Installation
|
Chapter 7
3
Cut the 1/8“ feed tube from the IC system at the reducing adapter on the
Dionex IC Pure system and connect both ends of the 1/8“ tube as
described in “Connection of a Dionex Integrion and Dionex ICS-2100,
Dionex ICS-4000, Dionex ICS-5000
+
, Dionex ICS-5000 or Dionex
ICS-3000system on page 35, Step 4 and 5.
4
Take the 1/8“ feed tube for the second Dionex Integrion and Dionex
ICS-2100, Dionex ICS-4000, Dionex ICS-5000
+
, Dionex ICS-5000 or
Dionex ICS-3000 system and connect it to the top connection on the T
piece.
5
Switch the Dionex IC Pure system back on and activate pure water
pumping for both Dionex Integrion and Dionex ICS-2100, Dionex
ICS-4000, Dionex ICS-5000
+
, Dionex ICS-5000 or Dionex ICS-3000
system using the Chromeleon CDS software.
Step
Action
Figure
Note
Again, ensure that the cutting ring and the taper
coupling are installed correctly (see “Connecting the control
cable from the IC Pure system to a Dionex Integrion and Dionex
ICS-2100, Dionex ICS-4000, Dionex ICS- 5000+, Dionex
ICS-5000 or Dionex ICS-3000 system” , Step 4)
.
1/8“ tube
T piece
Note
The system must be switched off.
The feedwater supply must be closed.
Check that the IC System is not in operating mode.
Connecting the Suppressor to the Dionex IC Pure system
Step
Action
Figure
1
Switch off the Dionex IC Pure system and stop pumping of pure water
for the Dionex Integrion Dionex ICS-2100, Dionex ICS-4000, Dionex
ICS-5000
+
, Dionex ICS-5000 or Dionex ICS-3000 system using the
Chromeleon CDS software.
2
Take the reducing adapter supplied with the Dionex IC Pure system and
carefully unscrew the union nut from the fitting.
Note
Ensure that the cutting ring and taper coupling that fit
inside the union nut are not lost when you unscrew the union
nut.
Reducing adapter
Union nut
Summary of Contents for dionex ic pure water
Page 10: ......
Page 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Page 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Page 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Page 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Page 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Page 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Page 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Page 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Page 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Page 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Page 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Page 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Page 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Page 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Page 127: ...Thermo Scientific IC Pure 117 ...