Installation
|
Chapter 7
6
Press the reducing adapter into the
“to pump“
connection on the
Dionex IC Pure system. When you hear an audible click you can be
sure that the quick-action fasteners have been inserted correctly.
7
Switch on the Dionex IC Pure system and the ion chromatogram.
As soon as the IC is switched on,
“Analyser“
appears on the control
display of the Dionex IC Pure system.
8
The Service Manager appears minimized as the Chromeleon Tray Icon
next to the clock in the Notification Area (also called System Tray) of the
Windows Taskbar. To open the Service Manager window, click (or
right-click) the Chromeleon Tray Icon.
9
If the Chromeleon Tray Icon is not present, you can start the Service
Manager by selecting Start> All Programs> Thermo Chromeleon 7>
Services Manager. Closing the Service Manager window minimizes it to
the Chromeleon Tray Icon.
10
In the Chromeleon CDS Server Configuration, open the corresponding
configuration settings and search for the entry “Dionex Integrion and
Dionex ICS-2100“ or “Dionex ICS-4000 system“ and right-click on the
entry you need.
Device Connection
Reducing adapter
Note
In the Analyser mode, the Nonstop and Interval modes
are not functional.
Note
If you are using an Dionex ICS-5000
+
, Dionex ICS-5000
or Dionex ICS-3000 system you must go to the tab 'Relays'
under the pump settings.
Summary of Contents for dionex ic pure water
Page 10: ......
Page 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Page 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Page 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Page 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Page 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Page 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Page 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Page 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Page 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Page 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Page 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Page 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Page 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Page 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Page 127: ...Thermo Scientific IC Pure 117 ...