Installation
|
Chapter 7
Calibrate suppressor regenerant water flow rate
7
Switch on the Dionex IC Pure system and start pumping of pure water
for the Dionex Integrion and Dionex ICS-2100, Dionex ICS-4000,
Dionex ICS-5000
+
, Dionex ICS-5000 or Dionex ICS-3000 system
using the Chromeleon CDS software.
8
Calibrate the suppressor regenerant water flow rate.
Step
Action
Figure
Note
Suppressor regenerant water flow rate must be adjusted using the flow control valve. The flow
rate should be approximately 3 to 5 mL/min for standard suppressors.
The flow rate should be determined with the suppressor connected to the suppressor supply
line of the Dionex IC Pure system and with the suppressor turned off.
The flow will be lower once the suppressor is switched on and should be checked again.
1
Record the weight of an empty container.
2
Insert the waste line from the suppressor into the container and weigh the contents after a known time interval.
3
Divide the weight of the contents of the vial (in grams) by the time to accumulate the sample. This will be the flow rate in
mL/min.
Note
If external water regeneration mode will not be used, the suppressor supply connection must
be closed off. Therefore, disconnect the reducing adapter from the ‘to Suppressor‘ connection on the
Dionex IC Pure system.
Note
A few minutes is sufficient.
Summary of Contents for dionex ic pure water
Page 10: ......
Page 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Page 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Page 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Page 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Page 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Page 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Page 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Page 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Page 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Page 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Page 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Page 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Page 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Page 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Page 127: ...Thermo Scientific IC Pure 117 ...