Initial start up
58
|
Dionex IC Pure system
Thermo Scientific
Chapter 10
Putting the system into operation
Note
Allow the system to warm up or to cool down to room temperature before starting it up for
the first time
.
CAUTION
Check that all tubing connections have been made as specified in the “Connections of
the Dionex IC Pure system” on page 22 section.
Switch the system on by pressing the “On/Off“ button. After a compulsory rinse, the system returns to
the last operating mode (Interval or Nonstop mode with IC system in Analyzer mode).
Note
If an IC System is connected to the unit and switched on, “Analyser“ will appear on the
control display of the Dionex IC Pure system. The Nonstop and Interval mode are then no longer
available. The user menu can continue to be used, however.
Note
To vent air from the system, switch the system 3 times successively to “Rinsing“ in the menu
and discard approx. 5 L of water each time with the manual dispenser. The ultra pure water limiting
value may come on during this process.
Use this 'NONSTOP'-key to switch the system to the “Nonstop“ operating mode.
When the unit has successfully reached the pure water value specified, return the unit to the
“Interval“ mode automatically or by pressing the “INTERVAL“ button while it is in the “Nonstop“
mode.
Summary of Contents for dionex ic pure water
Page 10: ......
Page 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Page 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Page 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Page 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Page 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Page 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Page 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Page 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Page 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Page 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Page 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Page 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Page 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Page 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Page 127: ...Thermo Scientific IC Pure 117 ...