Troubleshooting
100
|
Dionex IC Pure system
Thermo Scientific
Chapter 15
Dispensed water flow rate is too low
UF-Module blocked
Initial pressure too low (external
20 L tank with level control)
Internal pressure too low
Dispensed water flow rate out of
the tolerance range
0.2 µm final filter is clogged
Air look in unit
Replace UF-module (see
“Changing the ultrafilter” on
page 86)
Set the pressure regulator
Readjust pressure reducer
Contact Thermo Scientific for
calibration
Replace the 0.2 µm final filter
Dispense water at POU or rinse to
purge air from the system
Display of TOC content permanently
at 99 ppb
UV lamp not burned in / counter
reset
TOC measuring cell defective
Replace the UV lamp and reset the
operating hours counter (see
“Replacing the UV lamp” on
page 89)
Replace the TOC measuring cell
No analyzer is indicated at the
Dionex IC Pure system for operation
with an analyzer.
Settings in OEM menu are
incorrect
IC system not activated
Cable to IC system not
connected
Cable break
Correct the settings in the OEM
menu (see “OEM menu” on
page 72)
Activate IC system
Connect the cable to the IC system
Replace the cable
Wrong time or date
Time zone
Summer/winter time
Reset time and date
Wrong language
Wrong language set
Correct language setting
Error message: “Limit value feed“
Feedwater conductivity too high
Limiting value set too low
Check the feedwater measuring
cell and deactivate in the menu
Check and reset the limiting value
Display indIN
Measuring cell cable break
Replace measuring cell
Error message:“Pur.limit“
Ultrapure cartridge exhausted
Limiting value set too low
Replace ultrapure cartridge (see
“Changing the ultrapure cartridge”
on page 75
Check and reset the limiting value
Error message:“UV-time“
UV lamp operating time has
been exceeded
Replace the UV lamp (see
“Replacing the UV lamp” on
page 89)
Reset the operating time counter
Error
Cause
Remedy
Summary of Contents for dionex ic pure water
Page 10: ......
Page 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Page 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Page 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Page 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Page 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Page 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Page 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Page 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Page 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Page 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Page 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Page 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Page 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Page 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Page 127: ...Thermo Scientific IC Pure 117 ...