Set the limiting value for temperature
The maximum operating temperature limit for the system is set here. Should this temperature be
exceeded, the error message “Max. temperature” is triggered. This is shown in the 3rd line of the
display.
Basic setting: 35 °C
Setting range: 1 - 50 °C
After entering the OEM menu, press the “Menu“ button one time. The display shows:
Set the rinsing time
In this Menu point you can set the manually rinsing time. Rinse time is the amount of time the V4
Solenoid valve is opened to flush water to drain. This is helpful to purge air or impurities from the
system.
Basic setting: 30 sec.
Setting range: 10 - 60 sec.
After entering the OEM menu, press the “Menu“ button two times. The display shows:
Change the disinfection time
The disinfection time can be set in this menu. Disinfection time is the length of time required by the
instrument to circulate the cleaning solution in the system.
Basic setting: 30 minutes
Setting range: 15 - 90 minutes
After entering the OEM menu, press the “Menu“ button three times. The display shows:
Note
Use Menu button to scroll through menu items. Use enter only to save changed values.
Summary of Contents for dionex ic pure water
Page 10: ......
Page 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Page 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Page 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Page 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Page 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Page 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Page 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Page 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Page 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Page 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Page 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Page 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Page 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Page 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Page 127: ...Thermo Scientific IC Pure 117 ...