Maintenance
|
Chapter 13
Changing the ultrapure cartridge
Note
Replace the ultrapure cartridge when the “Lim. val.pure w.“ message is displayed indicating
the purity dropped below acceptable levels or when the “New filter set” message is shown in the
display.
Step
Action
Figure
1
Switch the system off.
2
Remove the cover in front of the ultrapure cartridge and press the
snapper and pull the cover towards the front to remove it.
3
Press on the two quick-connectors at the inlet and outlet of the
ultrapure cartridge and remove the used cartridge.
4
We recommend performing a disinfection when an existing ultra-
pure cartridge is replaced.
5
Place the ultrapure cartridge in the rear section of the unit and
insert the two quick-connectors into the ultrapure cartridge. When
you hear an audible click you can be sure that the quick-action
fasteners have been inserted correctly.
6
Put the cover for the ultrapure cartridge back in place and switch
the unit on.
Snapper
Cover
Note
Refer to section “Disinfection” on page 82
for an explanation of how to perform disinfection
Note
The quick-connectors are attached to the unit in such a
manner so as to prevent any confusion (switching).
Ultrapure cartridge
Summary of Contents for dionex ic pure water
Page 10: ......
Page 14: ... Transport and packaging 4 Dionex IC Pure system Thermo Scientific Chapter 1 ...
Page 18: ... Safety precautions 8 Dionex IC Pure system Thermo Scientific Chapter 2 ...
Page 22: ... Intended use of the device 12 IC Pure Thermo Scientific Chapter 4 ...
Page 28: ... Technical specifications 18 Dionex IC Pure system Thermo Scientific Chapter 5 ...
Page 62: ... Installation 52 Dionex IC Pure system Thermo Scientific Chapter 7 ...
Page 70: ... Initial start up 60 Dionex IC Pure system Thermo Scientific Chapter 10 ...
Page 74: ... Operating elements 64 IC Pure Thermo Scientific Chapter ...
Page 108: ... Waste disposal 98 Dionex IC Pure system Thermo Scientific Chapter 14 ...
Page 112: ... Troubleshooting 102 Dionex IC Pure system Thermo Scientific Chapter 15 ...
Page 116: ... Dionex IC Pure system spare parts 106 Dionex IC Pure system Thermo Scientific Chapter 16 ...
Page 118: ... Consumables 108 Dionex IC Pure system Thermo Scientific Chapter 17 ...
Page 119: ...Thermo Scientific Dionex IC Pure system 109 Terminal assignment ...
Page 120: ... Terminal assignment 110 Dionex IC Pure system Thermo Scientific Chapter 18 ...
Page 122: ... Maintenance record 112 Dionex IC Pure system Thermo Scientific Chapter 19 ...
Page 127: ...Thermo Scientific IC Pure 117 ...